Invasive Hemodynamic Ramp Testing in Pediatric VAD Support: Safety and Clinical Utility in a Bridge-to-Transplant Population
Authors
Affiliations
Abstract
INTRODUCTION
Invasive hemodynamic ramp testing is used to individualize ventricular assist device (VAD) support in adults, but pediatric experience remains limited. This is particularly relevant in children, in whom durable VAD support is predominantly used as a bridge to heart transplantation and physiologic targets may require greater individualization. We evaluated the safety, hemodynamic impact, and clinical utility of invasive ramp testing in pediatric VAD recipients.
METHODS
We performed a retrospective single-center study of patients <21 years supported with durable continuous-flow VADs who underwent invasive ramp testing from June 2018 through July 2025. Hemodynamics were obtained at baseline and after clinically directed pump-speed manipulation. Studies were categorized according to whether the final selected speed increased, decreased, or remained unchanged. Procedural safety, hemodynamic profiles, and subsequent clinical outcomes were assessed.
RESULTS
Fifteen patients underwent 19 completed ramp studies at a mean age of 13.4 ± 4.1 years and a median of 29 days (IQR 15, 62) after VAD implantation. Thirteen patients (87%) were supported as bridge to transplant. Eleven studies were performed for routine post-implantation or pre-discharge assessment, three for myocardial recovery, and five for cause, including low flow or elevated filling pressures. No major or minor adverse events attributable to the 19 completed ramp studies occurred. Final VAD speed differed from baseline after 10/19 studies (53%), increasing in 7. Nine studies (47%) resulted in retention of the baseline speed despite interrogation of alternative settings. Among studies resulting in a speed change, the mean percent change was 3 ± 2% of baseline speed. No measured invasive hemodynamic parameter changed significantly between baseline and final analyzed settings (all p > 0.05). Among 15 biventricular studies with complete CVP-PCWP data, only one changed hemodynamic profile category. Six patients were subsequently discharged home on VAD support a median of 11 days (IQR 2, 18) after ramp testing. Among 12 bridge-to-transplant patients who reached transplant listing, 11 (92%) underwent transplantation. One died on VAD support. Median time from ramp testing to transplantation was 66 days (IQR 34, 215).
CONCLUSION
Invasive hemodynamic ramp testing was safe in this predominantly bridge-to-transplant pediatric VAD population. However, nearly half of studies resulted in no change in pump speed, and when changes were made, they were minor adjustments, with only one patient who had a speed adjustment that resulted in an improved hemodynamic profile classification. These findings raise questions regarding the incremental value of routine invasive ramp testing in pediatric VAD recipients and suggest that a more selective, indication-driven approach may be appropriate. Larger multicenter studies are needed to identify patients most likely to benefit from invasive ramp assessment.